Teledyne 300 Carbon Monoxide Alarm User Manual


 
exiting calibration it is desirable to take a single instantaneous reading using the INST
mode. The table below summarizes the available sampling modes.
Sampling Modes
Mode Description
INST Records instantaneous reading
AVG Records average reading during reporting interval
MIN Records minimum reading during reporting interval
MAX Records maximum reading during reporting interval
Each parameter in this list can refer to a different (or the same) point, and can use a
different (or the same) sampling mode. For instance, you could create three parameters
which all measure the CO concentration for range 1 (the CONC1 parameter), but which
use the MIN, AVG, and MAX sampling modes. This permits you to record not only the
average concentration during the reporting interval, but the maximum excursions as well.
Like data channels, parameters also have properties which are listed in the table below.
Parameter Properties
Property Description Initial
Setting
Setting Range
PARAMETER The parameter to sample CONC1 Any of the parameters
listed in the table of
parameters shown
above
SAMPLE
MODE
The sampling mode to use
when reading this
parameter
AVG Any one of the INST,
AVG, MIN, and MAX.
PRECISION Number of decimal digits 1 0 - 3
The table below lists the data parameters initially defined. The entries in the Name and
Units columns are visible when editing and printing the data channel properties.
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 47