Teledyne 300 Carbon Monoxide Alarm User Manual


 
Property Setting DESCRIPTION
NAME “CO5MIN”
EVENT ATIMER Sample when automatic
timer is expired
PARAMETERS CONC1,AVG SO
2
Range #1 concentration
REPORT PERIOD 000:00:05 (i.e. 5 minutes) Average of 5 minutes
*
reported every 5 minutes
NUMBER OF
RECORDS
4032 (i.e. two weeks) Stores 4032 of Report
Period (5 minutes average)
RS-232 REPORT ON RS-232 is enabled
CHANNEL
ENABLED
ON Data channel is enabled
CAL. HOLD OFF ON Hold off is enabled
*
Sample measurement period of filtered CO concentration is 1 minute by default.
4.11.3 RS-232 reporting
Automatic RS-232 reporting can be independently enabled and disabled for each Data
Channel. For all default data channels, RS-232 reporting is initially set to “OFF.” If this
property is turned on, the Data Channel will issue a report with a time and date stamp to
the RS-232 port every time a data point is logged. The report format is shown below:
D 31:10:06 0412 CONC : AVG COCNC1=6.8 PPB
The report consists of the letter “D” follow by a time/date stamp(“31:10:06”) followed by
the instrument ID number(“0412”). Next is the Data Channel name(“CONC”) and the
sampling mode (“AVG” indicates that the data point is an average of more than one
sample as opposed to a instantaneous reading, “INST”). Finally, the name of the
parameter and its value(“COCNC1=6.8 PPB”) are printed. For Data Channels that
sample more than one parameter, such as PNUMTC and CALDAT, each parameter is
printed on a separate line.
To enable RS-232 reporting for a specific Data Channel:
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 49