System Operating Procedures
3-40 September 2000
Loading the Default Values
IMPORTANT: Record all processor settings before using this procedure.
Reloading the defaults causes the processor to be configured as
washless. The following settings and their calibrations return to
their original values.
• development time
• temperature settings
• replenishment rates
Press both Reset keys to save any changes to default settings and to return to
normal operation.
Changing the Wash/Stabilizer Setup
IMPORTANT: Record all setpoint and calibration data prior to using this
procedure.
1. Press and hold Set and both Reset keys. Release the Reset keys and then
release Set.
NOTE: The light above the left Reset keys illuminates.
2. Use the Up or Down Arrow to select the wash-stabilizer.
3. Press Set to access the screen in Set mode.
4. Use the Up or Down Arrow to change the operating setup.
5. Press Set to exit the Set mode.
6. To save the values, press both Reset keys simultaneously. Do not press Set.
NOTE: To make other changes, select the desired screen and enter new
values.
Changing the Metric/Fahrenheit Mode
IMPORTANT: Record all setpoint and calibration data for the processor prior to
using this procedure.
1. Press and hold Set and both Reset keys. Release the Reset keys and then
release Set.
2. Use the Up or Down Arrow to toggle to “Metricon”.
3. Press and hold the two black Reset keys.
4. When the processor begins to make a “beeping” sound, press the blue key.
The “beeping” sound will stop.
5. Press Set to reenter all setpoint and calibration data into the processor.
NOTE: Check that the temperature calibration is correct.