Kodak 20P Air Cleaner User Manual


 
System Operating Procedures
September 2000 3-35
Viewing/Setting the Temperatures
Viewing the Temperature Settings
1. Press Temperature on the OCP to access the Temperature screen.
2. Press the Up or Down Arrow to select the desired heater.
Setting the Temperature
1. View the temperature settings and select the desired heater.
2. Press Set.
3. Use the Up or Down Arrow to change the operating temperature.
4. Press Set to enter the Calibrate mode.
5. Press Set three times or press a different blue function key to return to View
mode.
NOTE: Press Temperature to toggle the heater between Manual Off and
Automatic mode.
Calibrating the Temperatures
IMPORTANT: Before calibrating a temperature sensor, make sure the
temperature has stabilized. The temperature near the probe is not
necessarily the same as the average temperature of the tank or
dryer.
1. View the temperature settings and select the heater.
2. Press Set twice to put the Temperature screen in Calibrate mode.
3. Use the Up or Down Arrows to enter an offset. Changing the offset causes
the displayed temperature to change. Adjust the offset until the displayed
temperature matches the actual temperature, as measured in the tank.
NOTE: Record and save the calibration value so it can be re-entered if the
memory is cleared.
4. Press Set twice or press a different blue function key to return to View mode.
Resetting the Min Max Temperatures
1. View the temperature settings and select the desired heater.
2. Press Set three times to enter the Min Max mode.
3. Press the Down Arrow to reset the values for the minimum and maximum
temperatures.
4. Press Set once or press a different blue function key to return to the View
mode.