Tektronix TDS520A Water System User Manual


 
Performance Tests
TDS 520A, 524A, 540A, & 544A Performance Verification
CONFIRM that the TRIG’D LED is lit and the waveform is dis-
played on screen.
c. Press SHIFT; then ACQUIRE MENU.
d. Press the main-menu Mode.
e. Select the side-menu Average. It should be already set to 16.
f. Press the main-menu Create Limit Test Template.
g. Press the side-menu V Limit.
h. Use the keypad to set V Limit to 100 mdiv (press 100, SHIFT, m, then
ENTER)
i. Press the side-menu OK Store Template.
j. Press MORE.
k. Press the main-menu Ref1.
l. Press CH1.
m. Press SHIFT; then ACQUIRE MENU.
n. Press the main-menu Limit Test Setup.
o. Toggle the side-menu Limit Test to ON.
p. Toggle the side-menu Ring Bell if Condition Met to ON.
q. Press the main-menu Mode.
r. Press the side-menu Envelope.
s. Use the keypad to set envelope to use 100 acquisitions (press 100,
then ENTER).
t. Press the main-menu Stop After button.
u. Press the side-menu Single Acquisition Sequence.
v. CONFIRM that the oscilloscope successfully makes 100 acquisitions.
If not successful, the oscilloscope bell will ring. When the word Run in
the top left corner of the display changes to STOP, the test is com-
plete (See Figure 1-29).