Tektronix TDS520A Water System User Manual


 
Performance Tests
Performance Verification Procedures
Third read the
results of the
DC
measurement here.
Second align the second
cursor to the DC test
level that you input.
First align a cursor to the
DC baseline (no input).
Figure 1-6: Measurement of the DC Accuracy for Delta Measurements
e.
Test all channels:
Repeat substeps a through d for all four channels.
3.
Reestablish the initial test hookup setup:
a.
Hook up the test-signal source:
Set the output of a DC calibration generator to 0 volts.
Move the BNC T connector back to CH 1.
b.
Initialize the oscilloscope:
Press save/recall SETUP.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
c.
Modify the default settings:
Press SHIFT; then ACQUIRE MENU.
Press the main-menu button Mode; then press the side-menu
button Average 16.
Press DISPLAY.
Press the main-menu button Graticule; then press the side-menu
button Frame.