Performance Tests
Performance Verification Procedures
Third read the
results of the
DC
measurement here.
Second align the second
cursor to the DC test
level that you input.
First align a cursor to the
DC baseline (no input).
Figure 1-6: Measurement of the DC Accuracy for Delta Measurements
e.
Test all channels:
Repeat substeps a through d for all four channels.
3.
Reestablish the initial test hookup setup:
a.
Hook up the test-signal source:
Set the output of a DC calibration generator to 0 volts.
Move the BNC T connector back to CH 1.
b.
Initialize the oscilloscope:
Press save/recall SETUP.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
c.
Modify the default settings:
Press SHIFT; then ACQUIRE MENU.
Press the main-menu button Mode; then press the side-menu
button Average 16.
Press DISPLAY.
Press the main-menu button Graticule; then press the side-menu
button Frame.