Bacharach ADM800 Smoke Alarm User Manual


 
84 Instruction 3015-4275
ADM800 - Specifications
Product Description: The ADM800 Ammonia Display Module provides remote programming, interrogation
and display functionality to support the AGM300 Ammonia Gas Monitor. The system
design supports compliance to the refrigerant monitoring requirements of ANS/BSR
ASHRE 15-1994.
Inputs: The ADM800 accepts inputs from up to four AGM300 monitors. It offers a wide variety
of displays and can fully program any associated monitor.
Display: Back lit LCD
Power Indication: Green LED glows when AC power is applied
Alarm Indication: Red LED glows and alarm condition is displayed
Fault Indication: Yellow LED glows and fault condition is displayed
Password Protection: The ADM800 can be password protected to prevent the unauthorized editing of setup
parameters without disturbing the ability to monitor system outputs
Power Safety Mode: Fully automatic system reset; all parameters maintained
Monitoring Distance: Up to 4500 feet from furthest AGM300
Communications: Full two-way communication with AGM300 monitors via RS-485 interface. A second
RS-485 interface provided for connection to a Building Management System.
Relays: Two SDDT programmable alarm contacts are provided (Rated 5 amp 120 VAC)
Operating Temp: 32 to 113 °F (0 to 45 °C)
Ambient Humidity: 5 to 90% (non-condensing)
Size/Weight: 11"H x 10"W x 3"D - 5 lbs.
AC Power: 120 to 230 VAC, 50/60 Hz, 21 Watts
Mounting: Wall mount
Certification: UL 61010-1 and CE Mark
Warranty: Two (2) Years from date of shipment
Altitude Limit: 2,000 Meters