Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-3: Warnings and Test Functions
A-23
Table A-3: T300/T300M and M300E/EM Test Functions, Revision L.8
TEST FUNCTION NAME MESSAGE TEXT DESCRIPTION
RANGE RANGE=50.0 PPM
3
CO RANGE=50.0 PPM
3, 7
D/A range in single or auto-range modes.
RANGE1 RANGE1=50.0 PPM
3
CO RANGE1=50.0 PPM
3, 7
D/A #1 range in dual range mode.
RANGE2 RANGE2=50.0 PPM
3
CO RANGE2=50.0 PPM
3, 7
D/A #2 range in dual range mode.
CO2RANGE CO2 RANGE=20 %
7
CO
2
range.
O2RANGE O2 RANGE=100 %
10
O
2
range.
STABILITY STABIL=0.0 PPM
3
CO STB=0.0 PPM
3, 7, 10
CO2 STB=0.0 %
7
O2 STB=0.0 %
10
Concentration stability (standard deviation
based on setting of STABIL_FREQ and
STABIL_SAMPLES).
RESPONSE
2
RSP=0.20(0.00) SEC
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
COMEAS CO MEAS=4125.0 MV Detector measure reading.
COREF CO REF=3750.0 MV Detector reference reading.
MRRATIO MR RATIO=1.100 Measure/reference ratio.
AUTOZERO
4, 5
AZERO RATIO=1.234
Measure/reference ratio during auto-
reference.
SAMPPRESS PRES=29.9 IN-HG-A Sample pressure.
PURGEPRESS
9
PURGE=7.5 PSIG Purge pressure
VACUUM
8
VAC=6.8 IN-HG-A Vacuum pressure.
SAMPFLOW
6
SAMP FL=751 CC/M Sample flow rate.
SAMPTEMP SAMPLE TEMP=26.8 C Sample temperature.
BENCHTEMP BENCH TEMP=48.1 C Bench temperature.
WHEELTEMP WHEEL TEMP=68.1 C Wheel temperature.
O2CELLTEMP
10
O2 CELL TEMP=50.2 C O
2
sensor cell temperature.
BOXTEMP BOX TEMP=26.8 C Internal box temperature.
BOXTEMP2
4
BOX TEMP2=29.6 C Internal box temperature #2.
OVENTEMP
11
OVEN TEMP=30.1 C Oven temperature
PHOTOTEMP PHT DRIVE=2500.0 MV Photometer temperature.
COSLOPE SLOPE=1.000
CO SLOPE=1.000
7
CO slope for current range, computed
during zero/span calibration.
COSLOPE1 SLOPE1=1.000
CO SLOPE1=1.000
7
CO slope for range #1 in dual range
mode, computed during zero/span
calibration.
COSLOPE2 SLOPE2=1.000
CO SLOPE2=1.000
7
CO slope for range #2 in dual range
mode, computed during zero/span
calibration.
COOFFSET OFFSET=0.000
CO OFFSET=0.000
7
CO offset for current range, computed
during zero/span calibration.
COOFFSET1 OFFSET1=0.000
CO OFFSET1=0.000
7
CO offset for range #1 in dual range
mode, computed during zero/span
calibration.
06864B DCN6314