Teledyne t320u Carbon Monoxide Alarm User Manual


 
APPENDIX A-2: Setup Variables For Serial I/O Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-16
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
800,
2000
13
1800
5,19
SAMP_FLOW_SET cc/m
Warnings:
640–960,
1400–2200
5,19
1500–2500
13
0–5000
Sample flow warning limits. Set
point is not used.
SAMP_FLOW_SLOPE — 1
4.5
5,19
0.001–100
Slope term to correct sample
flow rate.
VAC_SAMP_RATIO — 0.53,
0.61
13
0.1–2
Maximum vacuum pressure /
sample pressure ratio for valid
sample flow calculation.
7.5 PURGE_PRESS_SET PSIG
Warnings:
2.5–12.5
0–100
Purge pressure warning limits.
Set point is not used.
30 SAMP_TEMP_SET ºC
Warnings:
10.1–100
0–100
Sample temperature warning
limits. Set point is not used.
30 BOX_SET ºC
Warnings:
5–48
0–100
Internal box temperature warning
limits. Set point is not used.
30
46
19,23
BOX2_SET
5
,
OVEN_SET
19,23
ºC
Warnings:
25–35
41–51
19,23
0–100
Internal box temperature #2 /
oven set point and warning limits.
BOX2_CYCLE
5
,
OVEN_CYCLE
19,23
Seconds 10 0.5–30
Internal box temperature #2/oven
control cycle period.
BOX2_PROP
5
,
OVEN_PROP
19,23
1/ºC 1
0.5
19,23
0–100
Internal box temperature #2/oven
PID proportional coefficient.
Proportional band is the
reciprocal of this setting.
BOX2_INTEG
5
,
OVEN_INTEG
19,23
— 0.1
0.02
19,23
0–100
Internal box temperature #2/oven
PID integral coefficient.
BOX2_DERIV
5
,
OVEN_DERIV
19,23
— 0 0–100
Internal box temperature #2/oven
PID derivative coefficient.
BENCH_CYCLE Seconds 2
15
19,23
0.5–30
Optical bench temperature
control cycle period.
BENCH_PROP 1/ºC 5
1.5
19,23
0–100
100V optical bench temperature
PID proportional coefficient.
Proportional band is the
reciprocal of this setting.
BENCH_INTEG — 0.5
1.5
19,23
0–100
100V optical bench temperature
PID integral coefficient.
06864B DCN6314