Teledyne t320u Carbon Monoxide Alarm User Manual


 
Teledyne API – Model T300/T300M CO Analyzer Communications Setup and Operation
153
6.6. USB PORT (OPTION) FOR REMOTE ACCESS
The analyzer can be operated through a personal computer by downloading the TAPI
USB driver and directly connecting their respective USB ports.
1. Install the Teledyne T-Series USB driver on your computer, downloadable from the
Teledyne API website under Help Center>Software Downloads (www.teledyne-
api.com/software).
2. Run the installer file: “TAPIVCPInstaller.exe”
3. Connect the USB cable between the USB ports on your personal computer and your
analyzer. The USB cable should be a Type A – Type B cable, commonly used as a
USB printer cable.
4. Determine the Windows XP Com Port number that was automatically assigned to
the USB connection. (Start Control Panel System Hardware Device
Manager). This is the com port that should be set in the communications software,
such as APIcom or Hyperterminal.
Refer to the Quick Start (Direct Cable Connection) section of the Teledyne APIcom
Manual, PN 07463.
06864B DCN6314