HP (Hewlett-Packard) HP 83620A Vacuum Cleaner User Manual


 
Synthesized Step
Accuracy: Same as time base
Sweep
Minimum Step Size: Same as frequency resolution
Number of Points: 2 to 801
Switching Time: Same as CW
Dwell Time: 100 ps to 3.2
s
Synthesized List Mode
Accuracy: Same as time base
Minimum Step Size: Same as frequency resolution
Number of Points: 1 to 801
Switching Time: Same as CW
Dwell Time: 100
~LS
to 3.2
s
Ramp Sweep Mode
Accuracy2
(sweep time
2
100 ms and
5
5 s):
Sweep Widths
5
n x 10 MHz: 0.1% of sweep width
f
time base
accuracy.
Sweep Widths > n x 10 MHz: Lesser of 1% of sweep width or n x
1 MHz
+
0.1% of sweep width.
Sweep Time: 10 ms to 100 seconds, 300
MHz/ms
maximum rate
Internal 10 MHz
Accuracy: Calibration
f
Aging Rate
f
Temperature Effects
f
Line
Time Base
Voltage Effects
Stability
Aging Rate: 5 x
lo-lo/day,
1 x
10e7/year
With Temperature: 1 x
10-‘“/oC,
typical
With Line Voltage: 5 x 10
-lo
for line voltage change of
lo%,
typical
2
Sweeptime
1
150
ms
and
5
5 s for Option 006 instruments.
Specifications
2c-3