Peak RF Always
Function Group
POWER, USER CAL
Menu Map
5,9
Description
This
softkey
appears in two locations: the POWER Tracking Nenu
and the USER CAL Tracking Menu. The operation is the same in
both locations.
This
softkey
causes the synthesizer, when in CW or manual-sweep
output mode, to align the output filter (SYTM) so that its
passband
is centered on the RF output. Peaking is used to obtain both the
maximum available power and spectral purity, and the best pulse,
FM envelopes, at a given frequency. This peaking occurs each time
the frequency is changed, or every seven minutes. An asterisk next to
the key label indicates this function is active.
Programming Codes
SCPI: CALibration:PEAKing:AUTO
ON~OFF~l~O
Analyzer:
RPl
function on, RPO function off.
See Also
Auto Track, Peak RF Once, Tracking Menu
“Optimizing Synthesizer Performance,” in Chapter 1.
HP 8360
User’s Handbook
Operating and Programming Reference P-l