SWEep : STEP value is
(StopMAX-StartMIN)
/lO
SWEep:TIMEMINimum
SWEep:TIME:AUTO ON
SWEep:TIME:LLIMit
10ms
SWEep :
GENerat
ion
ANALog
SWEep : MODE AUTO
SWEep :
MANual
:
POINt
1
SWEep:MANual[:RELative]
0.50
SWEep:MARKer:STATe OFF
SYSTem:
ALTernate
1
SYSTem:ALTernate:STATe OFF
SYSTem:COMMunicate:GPIB:ADDRess 19
SYSTem:KEY:ENABle SAVE
SYSTem:MMHead:SELect:AUTO ON
SYSTem:SECurity:COUNt 1
UNIT:AMPCT *UNIT:POWerDBM
l
*SAV
<num>
The present instrument state is stored in the specified memory
register. The acceptable numeric range is from 1 to 8. An execution
error occurs if you try to save state 0.
l
*SRE
<num>
l
*SRE?
Sets and queries the value of the Service Request Enable Register.
l
*STB?
Queries the Status Byte. This is a non-destructive read.
l
*TRG
This command performs the same function as the Group Execute
Trigger command defined by IEEE 488.1.
l
*TST?
A full
selftest
is performed, without data logging or looping, and
returns one of the following error codes:
Error Code
Definition
0
Test passed.
1
Test failed.
2
Test not run yet. (This is an unlikely event.)
3
Test aborted.
4
Can not execute the test.
5
Can not execute the test, test skipped.
-1
Unrecognized result, software defect.
a
*WA1
This causes the synthesizer to wait for the pending commands to
be completed before executing any other commands. For example,
HP 8360
User’s Handbook
Operating and Programming Reference