SCPI
COMMAND SUMMARY
MARKer[n]:MODE How the frequency of the marker is determined.
FREQuency
Absolute frequency is used. The limits are
confined
to the present START and STOP
frequency limits.
DELTa
The value is specified with respect to the
reference marker. (MARKer[n]:REFerence)
The *RST values are the same as the FREQ :
CENTcommand
*RST
value.
l
MARKer
EnI
: MODE
FREQuency I
DELTa
l
MARKer
[n]
: MODE?
Sets and queries the mode of the specified marker. Setting one
marker to delta turns all other marker modes to frequency. If [n] is
not specified, the default is one. *RST value is FREQuency.
l
MARKer
[n]
: REFerence
<r-r>
l
MARKer
[nl
:
REFerence?
Sets and queries which marker is the reference marker for use in
the delta mode. While [n] may be used, there is really only a
single reference for all the markers.
MARKerl
:REFerence
5; and
MARKer2:REFerence
5; both set marker 5 as the reference.
l
MARKer
[n]
[
:
STATeI
ON | OFF | 1
|O
l
MARKer
[n]
[
:
STATe]
?
The state of the specified marker is set and queried (marker number
one if
[n]
is not specified). The *RST value for all markers is OFF.
l
MEASure
: AM?
A query-only command that causes the modulating AM signal to be
measured and the absolute value of the peak percent deviation to be
returned.
a
MEASure
: FM
A query-only command that causes the modulating FM signal level
to be measured and the corresponding peak frequency deviation
returned.
l
MODulation:OUTPut:SOURceAMIFM
l
MODulation:OUTPut:SOURce?
Sets and queries the source of the rear panel output modulation
BNC.
l
MODulation:OUTPut:STATe
ONlOFFlllO
l
MODulation:OUTPut:STATe?
Sets and queries the state of the rear panel output modulation BNC.
HP 8360
User’s Handbook
Operating and Programming Reference
S-41