888Xi Advanced Electronics Instruction Manual Section 2: Installation
PN 51-6888Xi December 2012
cable.
b. Pull the gland body away from the plastic insert. Use care not to damage the cable
shield braid.
c. Insert the cable wires into the proper entry port in either the probe housing or the
6888Xi enclosure.
d. At the probe housing, apply Teflon tape or similar sealing compound to the tapered
pipe threads. Thread the gland body into the probe housing until properly seated.
e. At the 6888Xi enclosure, insert the gland body into the left front cable port from the
inside of the enclosure. Use the rubber O-ring provided to seal the cable port.
f. Ensure the cable shield braid is evenly formed over the gray insert. When properly
formed, the braid should be evenly spaced around the circumference of the insert and
not extend beyond the narrow diameter portion.
g. Carefully press the gray insert into the gland body. The grooves on the insert should
align with similar grooves inside the gland body. Press the insert in until it bottoms out
in the gland body.
h. Slide the locking nut up and thread it onto the gland body. Tighten the locking nut so
the rubber grommet inside the plastic insert compresses against the cable wall to pro-
vide an environmental seal.
4. At the 6888Xi, connect the cable leads to the connectors on the transmitter I/O board as
indicated in Figure 2-7.