2.1 System Considerations
A typical system installation for a 6888Xi and O2 Probe is shown in Figure 2-1.
2.2 Mechanical Installation
6888Xi Advanced Electronics
The 6888Xi Advanced Electronics is available in a panel mounting, wall mounting, or pipe
mounting configuration. Refer to Figure 2-2 or Figure 2-3 for the panel, wall, or pipe mounting
details.
1. Ensure all components are available to install the 6888Xi.
2. Select a mounting location near or removed from the O2 Probe. Consider the temperature
limitations of the 6888Xi (see "Specifications") when selecting the mounting location.
3. Mount the 6888Xi at a height convenient for viewing and operating the interface.
Approximately 5 ft (1,5 m) is recommended.
4. The keypad window on the 6888Xi may have interior and exterior protective membranes.
Remove the protective membranes prior to use of the 6888Xi enclosure. Failure to remove
the protective membranes may cause the display to appear distorted. The membrane may
be difficult or impossible to remove after extended use at elevated temperatures.
12 Installation
S
ection 2: Installation 6888Xi Advanced Electronics Instruction Manual
December 2012 PN 51-6888Xi
FIGURE 2-1. Typical System installation