68 Using AMS Device Manager to configure the meter
Section 4: Configuring a 3818 LNG Meter Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual
May 2012 3-9000-771 Rev A
d. Click the All Variables tab to view a graphical display of the Primary, Secondary,
Third and Fourth Variables.
Figure 4-8 AMS Device Manager - Service Tools All Variables status indicators
9. Click the
OK button to return to the Overview page.
10. Enable the Write Protect switch on the CPU Module to protect the meter’s
configuration.
11. From the Overview window click
Display Meter K-Factors. K-Factors are a read-only
values calculated from the Full scale volumetric flow rate used with frequency outputs
and the Maximum frequency for frequency output.
Figure 4-9 Display Meter K-Factors
Click
Next to return to the Device Manager Overview page.