Emerson 3-9000-771A Carbon Monoxide Alarm User Manual


 
62 Using AMS Device Manager to configure the meter
Section 4: Configuring a 3818 LNG Meter Daniel 3818 LNG Liquid Ultrasonic Meter Installation Manual
May 2012 3-9000-771 Rev A
4.2 Using AMS Device Manager to configure the
meter
This procedure assumes you have AMS Device Manager installed on the host computer and have
downloaded the latest Daniel Liquid Ultrasonic Meter Device Description (DD).
If not installed, click the link below to download the AMS device installation tool kit.
http://www2.emersonprocess.com/en-US/documentation/deviceinstallkits/Pages/devicein-
stallkitsearch.aspx
Procedure - installing AMS Device Description
1. Use the link above to search for the Device Description (DD) for your Daniel 3810 Series
Liquid Ultrasonic Flow Meter.
2. Use the pull-down menu and select the Brand/Manufacturer -
Emerson Daniel Industries.
3. Next select the Device,
Liquid 3810 Series from the pull-down menu.
4. Choose the
Device Revision1, from the pull-down menu.
5. Next, select
HART from the Communication Protocol menu.
6. Select
AMS Device Manager for the Host System.
7. Select the
Host System Revision 11.5.
8. Verify your search parameters are correct, as shown below.
Figure 4-2 AMS Device Description search
9. Click the
Search Now button.