Teledyne GFC 7001EM Carbon Monoxide Alarm User Manual


 
Maintenance Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 261
Table 12-1: GFC 7001E/EM Maintenance Schedule
ITEM ACTION FREQ
CAL
CHECK
REQ’D
MANUAL
DATE PERFORMED
Particulate
Filter
Replace
Weekly or As
Needed
No
Verify Test
Functions
Record and
Analyze
Weekly or after
any
Maintenance or
Repair
No
Pump
Diaphragm
Replace Annually
Yes
Perform Flow
Check
Check Flow Annually
No
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance or
Repair
No
Pneumatic
lines
Examine and
Clean
As Needed
Yes if
cleaned
Cleaning Clean As Needed
Only if
cover
removed