Emerson NGA2000 PMD Home Security System User Manual


 
Instruction Manual
760007-A
July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-39
Model NGA2000 PMD
-- Analyzer Module Version Information --
Serial number:
Manufacturing date:
Hardware revision:
Software revision:
Revision date:
Revision time:
000.0
000.0
000.0
000.0
000.0
000.0
Measure
Back…
(C) Copyright Fisher-Rosemount Analytical Inc., 2000
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 /USA
Tel: (714) 986-7600
FAX: (714) 577-8739
Measure
More…
Back…
If it won’t calibrate…
Check that you are flowing the correct gas, and the
gas concentration is what it is supposed to be.
Make sure that the reading is stable before staring.
If you have enabled or disabled the linearizer, you
may have made it hard for the analyzer to calibrate.
If so, go to the calibration parameters screen under
Expert controls and set up, under Analyzer set up,
and disable the limits checking. Recalibrate, and then
enable the limits checking again.
If all else fails, manually adjust the calibration factors
HOME
ESCAPE INFO
Menu: 114 CALFAIL
Menu: 115 ABOUT
Menu: 116 ABOUT1