Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-23
Model NGA2000 PMD
Current measurement parameters help
Shows the main measurement parameters.
These can be controlled in the various
set up menus.
HOME
ESCAPE INFO
Barometric pressure parameters
Pressure transducer:
Barometric pressure compensation:
Measured pressure:
Transducer offset:
Transducer slope:
Transducer PGA gain:
000.0
000.0
000.0
000.0
000.0
000.0
HOME
ESCAPE INFO
Barometric pressure parameters
Shows whether the optional pressure transducer
is installed or not. Also this screen allows you to
set the transducer. Use the offset at atmospheric
pressure, and pull a vacuum of about 25cm H2O
on the transducer, and use the slope to make it read
correctly.
Choose whether to enable barometric compensation,
and whether to allow the analyzer to emit the
pressure value. Only one analyzer at a time in a
HOME
ESCAPE INFO
Menu: 66 MPARMSI1
Menu: 67 BAROM_PARMS
Menu: 68 BAROM_PARMI1