Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-27
Model NGA2000 PMD
Gas concentrations
Point 1
Gas value:
Raw reading:
Linearized value:
Point 2
Gas value:
Raw reading:
Linearized value:
Point to be measured:
000.0
000.0
000.0
000.0
000.0
000.0
000.0
HOME
ESCAPE INFO
DAT
A
NEXT
Polynomial set up
Select the range to linearizer.
Make sure that the span gas value is correct.
Choose whether to define the gas concentrations as
absolute values or as a percent to the span gas.
Use percent if you are diluting the span gas for this.
Get into the gas concentrations screens, and set the
concentration for as many points as you want.
At each point, flow the gas of the correct value, and
when the reading is stable, press
The analyzer will store the gas value and the reading
HOME
ESCAPE INFOMORE
Gas concentrations
Point 3
Gas value:
Raw reading:
Linearized value:
Point 4
Gas value:
Raw reading:
Linearized value:
Point to be measured:
000.0
000.0
000.0
000.0
000.0
000.0
000.0
HOME
ESCAPE INFO
DAT
A
NEXT
Menu: 78 POLYGAS1
Menu: 79 POLYSETI1
Menu: 80 POLYGAS2