Instruction Manual
760007-A
July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Model NGA2000 PMD
Range 1 (2, 3, 4) Factors. (Do steps 4
and 5 for each range.)
4. Select Zero Offset, press ENTER, adjust
the value to 525000 with the up/down ar-
row keys, and press ENTER. Do not
press ESCAPE at any time unless reten-
tion of prior settings is desired.
5. Select Span Factor, press ENTER, adjust
the value to 0.000015 with the up/down
arrow keys, and press ENTER. Do not
press ESCAPE unless retention of prior
settings is desired.
6. Attempt to recalibrate the Analyzer Mod-
ule according to the procedure outlined at
the beginning of Section 3-5 on page 3-5.
If re-calibration fails, return to the Range
Factors menu, readjust factors, and at-
tempt calibration again.