Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Model NGA2000 PMD
WARNING
OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply, adding Rosemount
Analytical PN 90331 Current Protector in series with the 24 V positive line will prevent over-voltage
spiking and resultant fuse flowing when powering up the instrument.
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping
front panel while hand or fingers are inside either handle can cause serious injury.
CAUTION
OVERBALANCE HAZARD
This Analyzer Module may tip instrument over if it is pulled out too far and the Platform is not prop-
erly supported.
NOTICE
Software compatibility is necessary for all NGA2000 components in your system to work together.
The version of your Platform's software must be equal to or greater that the version of any other
module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to
order software upgrade kit PN 657150 for the Platform.
You can locate the version of each NGA2000 component as follows:
Platform Controller Board
Turn power ON.
The display will show "Control Module V2. ...". This is the software version.
Analyzer Module
Located on the right side of the Analyzer Module case.
I/O Module
Located on the backplane connector of the module. If no label is present, the module is Version
2.0.