7-26 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Linearization functions
Polynomial set up…
Midpoint correction set up…
Use the polynomial set up to generate a linearizing
polynomial from up to 20 gases. With more than 6 gases
it will produce a fourth order polynomial linearizer.
Use the midpoint correction for a piecewise-linear final
correction, to bring up to three points precisely onto
HOME
ESCAPE INFO
Polynomial set up
Range to be linearized:
Current span gas:
Calculated polynomial order:
Gas values shown as:
Gas concentrations…
000.0
000.0
000.0
000.0
HOME
ESCAPE INFOCALC
Midpoint correction set up
Range 1
Correction:
Point being measured:
Point 1 gas concentration:
Point 2 gas concentration:
Point 3 gas concentration:
Point 1 reading:
Point 2 reading:
Point 3 reading:
Span gas value:
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
INFO
HOME
ESCAPE SET NEXT
Menu: 75 LINFUNCT
Menu: 76 POLYSETUP
Menu: 77 MIDPOINT1