Instruction Manual
760007-A
July 2003
7-38 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
nalyzer span
Are you sure?
You must have span gas flowing through the analyzer.
This control does NOT control any auto-calibration
module bound to this analyzer!
If you are sure, press SPAN again now.
Press the left arrow key when you are done.
Calibration status: 000.0
HOME
ESCAPE INFO
SPAN
A
nalyzer zero
Are you sure?
You must have zero gas flowing through the analyzer.
This control does NOT control any auto-calibration
module bound to this analyzer!
If you are sure, press ZERO again now.
Press the left arrow key when you are done.
Calibration status: 000.0
HOME
ESCAPE INFO
ZERO
Trend display control
The analyzer stores 24 hours of 15 minute averages.
These values are only accessible via a PC.
Use the variables DATA_INDEX and DATA_POINT
to access them.
HOME
ESCAPE INFO
Menu: 111 STOREDPVA
Menu: 112 ZERO_NOW
Menu: 113 SPAN_NOW