PERFORMANCE
Temperature control range
Ambient temperature+5 to 50
o
C (ambient temperature: 5 to 35
o
C)
Temperature distribution
± 0.25
o
C* (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, no load)
Temperature variation
± 0.1
o
C (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, no load)
CO
2
control range
0 to 20%
CO
2
variation
± 0.15% (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, no load)
Chamber humidity
95% ± 5%
Maximum heat emission
1,120 kJ/h
Applicable environment
condition
Temperature: 5 to 35
o
C, Humidity: 80% max.
(The designed performance may not be obtained if
the ambient temperature is equal or less than 15
o
C)
Noise level
30 dB (A scale)
Maximum power consumption
350 W 310 W
110 to 120 V, 60 Hz 220 to 240 V, 50 Hz
Total maximum
current
2.8 A 1.4 A
Note: 㪫㪿㪼㩷㫌㫅㫀㫋㩷㫎㫀㫋㪿㩷㪚㪜㩷㫄㪸㫉㫂㩷㪺㫆㫄㫇㫃㫀㪼㫊㩷㫎㫀㫋㪿㩷㪜㪚㩷㪻㫀㫉㪼㪺㫋㫀㫍㪼㫊㪅㩷
* Based on our measuring method.