Sanyo MCO-19AIC Carbon Monoxide Alarm User Manual


 
134
43
6. Press the H
2
O
2
Key for 5 seconds. The system check will start.
7. If the system is normal, the following display will appear. (If the system is not normal, refer to Table 2
on page 47.) Select OK and press the ENTER Key to start H
2
O
2
decontamination. Decontamination will
be performed automatically to step 10. The buzzer will sound when decontamination has been completed
(It takes about 1 hour 40 minutes after warming-up).
Note: The outer door will be locked with an electric lock for safety until the decontamination has been
completed.
8. During the H
2
O
2
mist generation, “H2O2 Decon” will flash at the top left corner of the screen.
Note: H
2
O
2
mist generates from the H
2
O
2
Generator and then quickly changes to a gas.