AUTOMATIC CO
2
DENSITY CALIBRATION
㩷
To calibrate the CO
2
density, an Automatic CO
2
Standard Gas Calibration Kit (MCO-SG) is available. By
connecting a CO
2
standard gas cylinder that has a density that is the same as the set CO
2
density (usually
5% CO
2
), the CO
2
sensor can be calibrated to maintain the accuracy of the CO
2
density display as
necessary.
㩷
Ɣ Setting connected standard CO
2
gas density
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
2. Press the Down Cursor Key from the Select Tools (1/2) Display to display the Select Tools (2/2) Display,
and then select Standard Gas Setting, press the MENU Key to display the menu, select OK, and press
the ENTER Key.