133
42
H
2
O
2
decontamination
Use the following procedure to perform H
2
O
2
decontamination using the H
2
O
2
Generator (MCO-HP).
1. Take out all trays, tray supports, side supports, the humidifying pan cover, the humidifying pan, the fan
cover, and the duct from the chamber. Dispose of the water in the humidifying pan, wipe the inside walls
of the chamber with a gauze containing water or alcohol for sterilization.
2. Attach the duct and side supports, and set the tray supports in the 2nd, 5th, 7th, 9th holes from the top
of the side supports. Then set the trays.
3. Pour one bottle of H
2
O
2
Decontamination Reagent (MCO-H2O2) into the H
2
O
2
Generator. Set the two
pins on the H
2
O
2
Generator in the 2 holes on the lower left side of the duct (Fig. 1).
4. Connect the H
2
O
2
Generator and connector on the bottom right of the far side of the chamber with the
enclosed cable. Be sure to keep the connector cap (Fig. 1).
5. Set the humidifying pan cover, humidifying pan, and fan cover as shown in Fig. 2, and close the inner
door and outer door (Fig. 2).
Note: The trays included as accessory are designed to be appropriate for decontamination. If Half Trays
(MCO-25ST, optional) or Trays for previous models are used, decontamination may not be sufficiently
effective.
H
2
O
2
Generator
Fig. 2
Humidifying pan
Humidifying pan cover
Fan cover
Front
Fig. 1
H
2
O
2
Generator
UV lamp
Connector
Ultrasound oscillator
Front
Front