Sanyo MCO-19AIC Carbon Monoxide Alarm User Manual


 
119
28
Setting the UV lamp ON period
Use the following procedure to change the setting of the UV lamp ON period.
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
2. Select UV Setting from the Select Tools Display, press the MENU Key to display the menu, select OK,
and press the ENTER Key.
3. The UV Setting Display will appear. Set the UV Timer for the UV lamp ON period.
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key. The parameter will
be saved.
1
T e
m p 3 7 . 0
o
C C O 2 5 . 0 %
2
3
4
5
S t a t u s
6
O K
7
㧟㧣㧜
㧡㧜
M E N U
S e t
L o g
T o o l s
S e l e c t T o o l s (1/2)
T e m p / C O2 C a l i bration
A l a r m Se t t i n g
L C D / D A Q S e t t i n g
U V S e t t i n g
K e y L o ck P W S etting
D a t e T i m e
M E N U
O K
S v c
C a n c e l
1
U V S e t ti n g
3
U V T i m e r
5 mi n ( 0 - 3 0 m i n )
U V L i f e 0 %
4
U V T i m e r E x t . + 0 %
5
U V 2 4 h M o d e 0 (0.OFF 1.ON )
6
M E N U
O K
C a n c e l