Chapter 6 NI-VXI Configuration Utility
VXI/VME-PCI8022 for Solaris 6-30
©
National Instruments Corporation
MXI CLK10 Signal
The VXI-MXI-2 can either receive or drive the MXIbus CLK10 signal.
In its default setting, the VXI-MXI-2 uses the switch setting of S7 to
determine the signal direction.
VME users—This option is not applicable to the VME-MXI-2.
You can use the Drive or Receive options to override the setting of S7
and control the direction of the MXIbus CLK10 signal. When receiving
the MXIbus CLK10 signal, configure the W3 jumper setting to use the
MXIbus as the source for generating the VXIbus CLK10 (applicable
only if the VXI-MXI-2 is a Slot 0 device). When driving the MXIbus
CLK10, the VXIbus CLK10 is used as the source. In this case, change
the jumper setting so that it does not use the MXIbus CLK10 as the
source for the VXIbus CLK10.
Caution: Do not configure more than one MXIbus device to drive MXI CLK10.
Setting up a second device to drive MXI CLK10 could damage the device.
!