Appendix D: ELECTROMAGNETIC COMPATIBILITY
D.1 COMPLIANCE AND WARNINGS
The MR810 with RT308 breathing circuit has been tested to and found compliant with the requirements of
IEC 60601-1-2: 2001.
WARNING:
The use of other accessories other than those specified may result in increased
emissions or decreased immunity of the equipment.
Verify correct operation if the MR810 is to be used adjacent to other equipment.
D.2 ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The MR810 is intended for use in the electromagnetic env ironment specified below. The customer or the user
of the MR810 should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The MR810 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and not likely to cause any interference in nearby
electronic equipment.
RF emissions Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The MR810 is suitable for use in all establishments
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.