Fisher & Paykel MR810 Humidifier User Manual


 
MR810 Technical Manual Revision C REF 185042601
27
6
6.5.7 INSTALLING THE HEATER PLATE
Figure 6.9: Heater plate assembly
1. Ensure the heater plate springs are located in their res pective location holes on the heater plate.
2. Pull the heater plate harness through the hole in the case top (Figure 13).
3. Carefully orient the case upper over the heater plate springs, then screw the three mounting screws
back into place.
6.6 REPLACING THE POWER CABLE OR HEATER-WIRE ADAPTOR
Note: The spare heater-wire adaptor, part number 043 042 322, includes the cable clamp and collet for the
mains cable and heater-wire adaptor.
Table 10: Part numbers and colour codes of Fisher & Paykel
Healthcare power cables
Model Supply
Voltage
Power Cable
Part Number
Phase
Colour
Neutral
Colour
Earth
Colour
Plug
MR810ALU
MR810AGU
MR810AFU
MR810ARU
MR810ANU
230 V~ 095 4 28 32 3 Brown Blue Green / Yellow Schuko
MR810AEA 230 V~ 095 4 28 31 7 Brown Blue Green / Yellow Australa si an
MR810AEK 230 V~ 09 5 4 28 56 9 Brown Blue Green / Yellow UK
MR810AEU 230 V~ 095 4 28 85 6 Brown Blue Green / Yellow No Plug
MR810AD U 230 V~ 095 4 28 85 4 Brown Blue Green / Yellow Dani sh
MR810JHU
MR810JSU
115 V~ 095 428 322 Brown or
Black
Blue or
White
Green / Yellow USA
Right angle
MR810GJU 100 V~ 095 428 869 Brown or
Black
Blue or
White
Green/Yellow Japanese
Heater plate
spring