Bacharach H25-IR Carbon Monoxide Alarm User Manual


 
Maintenance H25-IR
44 Instruction 3015-4342
7. From the C:\RFU folder, run the flash memory program by
double clicking the file rfu.exe.
8. Select Setup > Communications.
9. Under “Serial Options”, select Use Serial Connection. Then
set the Baud Rate to 115200, and select Enable Processor
Detection.
Note that COM1 is the default Comm Port. If necessary,
change the Comm Port to match the port that the instrument
was connected to in Step
2.
Click OK.
10. Select Setup > File Locations and verify that the coldload.bin,
pilot.bin, and flash.ini files are all located in the C:\RFU
folder.
Click OK if these files are in the correct folder. If necessary,
use the browse buttons in the Choose File Locations dialog
box to locate each of these files.