Tektronix TDS540A Water System User Manual


 
Performance Tests
Performance Verification Procedures
Press the side-menu Field Rate.
Use the keypad to set the field rate to 60.05 Hz (press 60.05,
then ENTER).
Press the side-menu Lines.
Use the keypad to set the field rate to 793 lines (press 793, then
ENTER).
Press the side-menu Fields.
Use the keypad to set the number of fields to 1 (press 1, then
ENTER).
Press the side-menu Sync Width.
Use the keypad to set the width to 400 ns (press 400, SHIFT, n,
then ENTER).
Press the side-menu –more– 1 of 2. Then press V1 Start Time.
Use the keypad to set V1 start time to 10.10 s (press 10.10,
SHIFT, , then ENTER).
Press the side-menu V1 Stop Time.
Use the keypad to set V1 stop time to 10.50 s (press 10.50,
SHIFT, , then ENTER).
Press the main-menu Type pop-up to select Edge.
Press HORIZONTAL MENU.
Press the main-menu Record Length.
Select the side-menu 1000 points in 20div. (if needed, first
press the side-menu –more– 2 of 2).
Turn the HORIZONTAL POSITION knob to position the trigger ‘T’
two divisions to the left of the center screen.
Press MEASURE.
If needed, press the main-menu Select Measrmnt.
Press the side-menu Negative Width.
Press the side-menu Period.
b. Setup the pulse generator (PG502) for Sync Duty Cycle Test
Set PULSE DURATION to 50 ns.
Set PERIOD to 10 s.
Set OUTPUT (VOLTS) to –1 for LOW LEVEL and +1 for HIGH
LEVEL.
Depress the COMPLEMENT button.
Be sure BACK TERM is depressed (in).