Omega Vehicle Security OM5-C Electric Heater User Manual


 
2
SCM5B
OMEGA® OM5 Series Isolated Signal Conditioning Modules are low cost,
high performance, plug-in signal conditioners. Each OM5 module provides
a single channel of isolated analog input or output. Input modules interface
to all types of external sensors. The modules filter, isolate, amplify, and
convert the input signal to a high-level analog voltage output. The output
modules accept a high-level analog voltage signal from a host system, then
buffer, isolate, and amplify before providing a process current or voltage
output to field devices. Over 115 different OM5 modules are available
encompassing a wide selection of isolated analog input and output functions.
Analog inputs include voltage and current in narrow and wide bandwidths,
Thermocouple, RTD, Potentiometer, Strain Gage, Frequency, and 2-Wire
Transmitter. Custom I/O ranges are also available. All modules are CSA
certified and FM approved for safe operation in Class I, Division 2, Groups
A, B, C, and D hazardous environments.
Accessories include: addressable and non-addressable single, dual, eight
and 16 channel backpanels, available with on-board temperature sensors
and cold junction thermocouple compensation; power supplies; mounting
racks; interface cables, and evaluation boards.
OM5 modules offer several advantages when compared with competitive
parts, while maintaining equivalent price:
50 times better noise rejection by using a 6-pole filter with 95dB
NMR, versus a three-pole filter with 60dB NMR; • Lower output noise
• True 3-way isolation • CMR of noise spikes measures 20dB better
than competing models.
Key Specifications
1500V Isolation
Accuracy, 0.05%
CMR, 160dB
NMR, 95dB
Transient Protection, ANSI/IEEE C37.90.1-1989
±1µV/°C Drift
Output noise as low as 150µVrms
240VAC Protection for field I/O
Operating temp range, –40°C/+85°C
CSA CERTIFIED (Class
I
, Division 2, Groups A, B, C, D)
FM APPROVED (Class
I
, Division 2, Groups A, B, C, D)
EUROPEAN EMC DIRECTIVE COMPLIANT
• CE APPROVAL for low voltage directive not applicable.
Products comply with ENG1010-1 (IEC1010)
Applications
• DESIGNED FOR INDUSTRIAL PLANT ENVIRONMENTS
• PROTECTS USER EQUIPMENT FROM LIGHTNING AND HEAVY
EQUIPMENT POWER-LINE VOLTAGE
• REDUCES ELECTRICAL NOISE IN MEASURED SIGNALS
• CONVENIENT SYSTEM EXPANSION AND REPAIR
General Description
ANALOG VOLTAGE INPUT MODULES, NARROW BANDWIDTH (4Hz BW)
MODEL INPUT RANGE OUTPUT RANGE*
OM5-IMV-10A-C ±10mV ±5V
OM5-IMV-50A-C ±50mV ±5V
OM5-IMV-100A-C±100mV ±5V
OM5-IMV-10B-C ±10mV 0 to +5V
OM5-IMV-50B-C ±50mV 0 to +5V
OM5-IMV-100B-C±100mV 0 to +5V
OM5-IV-1A-C ±1V ±5V
OM5-IV-5A-C ±5V ±5V
OM5-IV-10A-C ±10V ±5V
OM5-IV-1B-C ±1V 0 to +5V
OM5-IV-5B-C ±5V 0 to +5V
OM5-IV-10B-C ±10V 0 to +5V
OM5-IV-20A-C ±20V ±5V
OM5-IV-20B-C ±20V 0 to +5V
OM5-IV-40A-C ±40V ±5V
OM5-IV-40B-C ±40V 0 to +5V
ANALOG CURRENT INPUT MODULES, 4Hz AND 1kHz BANDWIDTH
MODEL INPUT RANGE OUTPUT RANGE* BW
OM5-II-4/20-C 4 to 20mA 0 to +5V 4Hz
OM5-II-0/20-C 0 to 20mA 0 to +5V 4Hz
TRUE RMS INPUT MODULES
MODEL INPUT RANGE (rms) OUTPUT RANGE (dc)
OM5-IAC-100B-C 0 to 100mV 0 to +5V
OM5-IAC-1B-C 0 to 1V 0 to +5V
OM5-IAC-10B-C 0 to 10V 0 to +5V
LINEARIZED 2- OR 3-WIRE RTD INPUT MODULES (0 to +5V OUTPUT*,
4Hz BW)
MODEL TYPE*** INPUT RANGE
OM5-IP-N100-C 100 Pt –100°C to +100°C (–148°F to +212°F)
OM5-IP-100-C 100 Pt 0°C to +100°C (+32°F to +212°F)
OM5-IP-200-C 100 Pt 0°C to +200°C (+32°F to +392°F)
OM5-IP-600-C 100 Pt 0°C to +600°C (+32°F to +1112°F)
OM5-IC-120-01-C 10 Cu at 0°C0°C to +120°C (+32°F to +248°F)
OM5-IC-120-02-C 10 Cu at 25°C0°C to +120°C (+32°F to +248°F)
OM5-IN-300-C 120 Ni 0°C to +300°C (+32°F to +572°F)
LINEARIZED 4-WIRE RTD INPUT MODULES (0 to +5V OUTPUT*, 4Hz BW)
MODEL TYPE*** INPUT RANGE
OM5-IP4-N100-C 100 Pt –100°C to +100°C (–148°F to +212°F)
OM5-IP4-100-C 100 Pt 0°C to +100°C (+32°F to +212°F)
OM5-IP4-200-C 100 Pt 0°C to +200°C (+32°F to +392°F)
OM5-IP4-600-C 100 Pt 0°C to +600°C (+32°F to +1112°F)
OM5-IC4-120-01-C 10 Cu at 0°C0°C to +120°C (+32°F to +248°F)
OM5-IC4-120-02-C 10 Cu at 25°C0°C to +120°C (+32°F to +248°F)
OM5-IN4-300-C 120 Ni 0°C to +300°C (+32°F to +572°F)
POTENTIOMETER INPUT MODULES (4Hz BW)
MODEL INPUT RANGE OUTPUT RANGE*
OM5-PT-100-C 0 to 100 0 to +5V
OM5-PT-500-C 0 to 500 0 to +5V
OM5-PT-1K-C 0 to 1K 0 to +5V
OM5-PT-10K-C 0 to 10K 0 to +5V
THERMOCOUPLE INPUT MODULES (0 to +5V OUTPUT*, 4Hz BW)
MODEL TYPE** INPUT RANGE
OM5-ITC-J-C J –100°C to +760°C (–148°F to +1400°F)
OM5-ITC-K-C K
–100°C to +1350°C (–148°F to +2462°F)
OM5-ITC-T-C T –100°C to +400°C (–148°F to +752°F)
OM5-ITC-E-C E 0°C to +900°C (+32°F to +1652°F)
OM5-ITC-R-C R 0°C to +1750°C (+32°F to +3182°F)
OM5-ITC-S-C S 0°C to +1750°C (+32°F to +3182°F)
OM5-ITC-B-C B 0°C to +1800°C (+32°F to +3272°F)
OM5-ITC-C1-C C +350°C to +1300°C (+662°F to +2372°F)
OM5-ITC-N1-C N -100°C to +1300°C (-148°F to +2372°F)
STRAIN GAGE INPUT MODULES (
±5V OUTPUT*, 10kHz BW)
MODEL INPUT EXCITATION
OM5-WBS38-01-C ±10mV Full Bridge Input, (3mV/V) 100 to 10k 3.333V
OM5-WBS-1-C ±30mV Full Bridge Input, (3mV/V) 300 to 10K 10.000V
OM5-WBS38-03-C ±10mV Half Bridge Input, (3mV/V) 100 to 10k 3.333V
OM5-WBS-3-C ±30mV Half Bridge Input, (3mV/V) 300 to 10K 10.000V
OM5-WBS-2-C ±20mV Full Bridge Input, (2mV/V) 300 to 10K 10.000V
OM5-WBS-4-C ±33.3mV Full Bridge Input, (10mV/V) 100 to 10K 3.333V
OM5-WBS-5-C ±100mV Full Bridge Input, (10mV/V) 300 to 10K 10.000V
ANALOG CURRENT OUTPUT MODULES, 400Hz AND 1kHz BANDWIDTH
MODEL INPUT RANGE OUTPUT RANGE BW
OM5-IVI-B4-C 0 to +5V 4 to 20mA 400Hz
OM5-IVI-A4-C ±5V 4 to 20mA 400Hz
OM5-IVI-B0-C 0 to +5V 0 to 20mA 400Hz
OM5-IVI-A0-C ±5V 0 to 20mA 400Hz
SELECTION GUIDE FOR OM5 ISOLATED