Energizer PORTABLE GAS DETECTION INSTRUMENT Smoke Alarm User Manual


 
42
PHOSPHINE SPECIFICATIONS
Range:
0.00 to 9.99 ppm
Response Time:
T90 < 30 seconds
Operating Temperature Range:
-20 to +50
o
C
(-4 to +122
o
F)
Graphic Display Temperature Range:
-10 to +60
o
C
(+14 to +140
o
F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20
o
C
(32 to 68
o
F)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Arsine at 150 ppb: 100 ppb
Silane, 1000 ppb: 900 ppb
Diborane, 300 ppb: 105 ppb
Germane Dioxide, 600 ppb: 550 ppb
Sulphur Dioxide, 5 ppm: 1 ppm
Hydrogen, 1000 ppm: 1 ppm
Ethylene, 100 ppm: 1 ppm
Carbon Monoxide, 1000 ppm: 5 ppm
Consult Factory for additional information.