6.9 CONFIGURATION MENU
The CONFIGURATION menu of the SCOTT INSTRUMENTS MINI-SA pro-
vides the user with the following two (2) optional features:
BATTERY SAVER mode can extend operational battery life to sev-
eral hundred hours by turning off the instrument graphic display dur-
ing normal non-alarm operation.
PASSWORD PROTECTION menu can be used to lock out all menus
except ZERO CALIBRATION for toxic gas sensors or AUTO CALI-
BRATION for an oxygen sensor, preventing unauthorized users from
changing instrument settings.
6.9.1 BATTERY SAVER MODE
The BATTERY SAVER mode extends the operational life of the batteries by
turning off the graphic display when it is not needed. To use this feature,
depress the “SELECT” pushbutton when the CONFIGURATION menu screen
is displayed as described in section 6.2. The following screen will appear:
CONFIGURATION
BATTERY PASSWORD
SAVER MENU
Depress the “BATTERY SAVER” pushbutton and the following screen will
appear:
CONFIGURATION
ENABLE BATTERY
SAVER MODE?
YES NO
Depress the “YES” pushbutton to enable the BATTERY SAVER mode. When
the BATTERY SAVER mode is enabled, the MINI-SA will beep once and the
Operating Mode screen will appear. To the right of the the battery level
indicator three (3) vertical plus signs (+) will appear, which signify that the
BATTERY SAVER mode is enabled. After thirty (30) seconds, the instrument
will beep once and the graphic display will go blank. While the graphic dis-
play is blank, the Green Status LED will flash at an increased rate of once
every five (5) seconds.
While the BATTERY SAVER mode is on, depressing either pushbutton will
turn on the graphic display for thirty (30) seconds.