Energizer PORTABLE GAS DETECTION INSTRUMENT Smoke Alarm User Manual


 
37
7.0 ERROR MESSAGES
WARNING
A SCOTT INSTRUMENTS MINI-SA INSTRUMENT THAT DISPLAYS
ANY ERROR MESSAGE MUST BE REMOVED FROM SERVICE IM-
MEDIATELY AND THE FAILURE DIAGNOSED AND REPAIRED BY
AUTHORIZED PERSONNEL. DO NOT ASSUME THAT A PROBLEM
HAS CORRECTED ITSELF IF A PREVIOUSLY DISPLAYED ERROR
MESSAGE DOES NOT APPEAR A SECOND TIME.
During calibration, if the SCOTT INSTRUMENTS MINI-SA detects a sensor
with insufficient output to be calibrated, the following screen will appear:
<gas> SENSOR
For Toxic Gases SPAN CALIBRATION
FAILURE
TURN OFF RECAL
OXYGEN SENSOR
For O
2
SPAN CALIBRATION
FAILURE
TURN OFF RECAL
If one of these screens appears, verify that the proper gas is flowing to the
instrument and all fittings are properly connected. Then press “RECAL” to
attempt the calibration again. If calibration is not successful, press “TURN
OFF” and service the instrument as needed.
On subsequent power on, the INITIALIZING line on the first screen
displayed (SCOTT INSTRUMENTS MINI-SA...) will be replaced with
CALIBRATION REQUIRED, until the instrument is successfully
recalibrated. This may require replacing the sensor.