If an alarm condition occurs while in the BATTERY SAVER mode, the graphic
display will come on automatically and remain on until the alarm is reset.
After the alarm indication clears, the graphic display will remain on for
thirty (30) seconds and then the display will go blank as described above.
If the BATTERY SAVER mode had already been enabled , the graphic display
will show the following:
DISABLE BATTERY
SAVER MODE?
YES NO
Depressing the “YES” pushbutton will turn off the BATTERY SAVER mode.
6.9.2 PASSWORD PROTECTION
The PASSWORD PROTECTION menu is intended to be used by authorized per-
sonnel for the purpose of locking out the menus to prevent changes to the instrument
settings. A four digit numeric password can be set up to restrict access to all menus
except AUTO CALIBRATION or ZERO CALIBRATION. To use this feature, de-
press the “SELECT” pushbutton when the CONFIGURATION MENU screen is dis-
played as described in section 6.2. The following screen will appear:
CONFIGURATION
BATTERY PASSWORD
SAVER MENU
Depress the “PASSWORD MENU” pushbutton and the following screen will
appear:
ENTER PASSWORD
0000
UP SET
Use the “UP” and “SET” push-buttons to enter the correct password. (The
MINI-SA comes from the factory with the default password: 1234.) Entering
an incorrect password will display an INCORRECT PASSWORD message
and allow the user to enter the password again. When the correct password