9.0 SPECIFICATIONS
HYDROGEN SULFIDE SPECIFICATIONS
Range:
0 to 100 ppm (for H
2
S 100)
0 to 500 ppm (for H
2
S 500)
Response Time:
T90 < 30 seconds
Operating Temperature Range:
-40 to +50
o
C
(-40 to +122
o
F)
Graphic Display Temperature Range:
-10 to +60
o
C
(+14 to +140
o
F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20
o
C
(32 to 68
o
F)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: < 2 ppm
Sulfur Dioxide, 5 ppm: - 1 ppm
Nitric Oxide, 35 ppm: < 0.7 ppm
Hydrogen, 10000 ppm: < 10 ppm
Nitrogen Dioxide, 5 ppm: - 1 ppm
Consult Factory for additional information.