28
Output setpoints
Simulated tests
Configure
Exit Enter
MODEL 54eA SECTION 5.0
SOFTWARE CONFIGURATION
5.5 CHOOSING DISPLAY OPTIONS
3. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
4. Move the cursor to "Configure" and press Enter (F4).
5. With the cursor on "Display", press Enter (F4).
6. A screen showing the present main measurement will appear. To
change the measurement, for example, from oxygen to ozone, press
Edit (F4), then use the key to scroll through the choices. Press
Save (F4) to store the setting.
A screen will appear warning that if the measurement is changed, the
analyzer will return to factory default settings. Press Cont (F3) to con-
tinue. Press Abort (F1) to cancel the change.
7. If oxygen is being measured, the screen at left appears. Use the
key to move the cursor to "Sensor" and press Edit (F4). Use the or
keys to scroll through the choices. Press Save (F4) to store the
selection.
Use the key to move the cursor to "Meas units" and press Edit (F4).
Use the or
keys to scroll through the choices. Press Save (F4)
to store the selection.
Display
Outputs
Alarms
Exit Enter
Measure
: Oxygen
Sensor: RMT Standard
Meas units: ppm
Exit Edit
Measure : Oxygen
Sensor: RMT Standard
Meas units: ppm
Exit Enter
1. The 54eA controller can be used with most amperometric sensors manufactured by Rosemount Analytical. The
user must configure the analyzer to match the sensor being used.
2. The display menu also lets the user customize the third line in the display, change timeout values, choose a lan-
guage other than English, and change the display contrast.
RMT Standard 499ADO or 499ATrDO
RMT Biopharm Hx438 or Gx448
SSDO Other Steam sterilizable sensor from other manufacturers
ppm (mg/L) use with any sensor except 499ATrDO
ppb (ug/L) use with 499ATrDO sensor
% sat (concentration as a percent of saturation value)
use with any sensor except 499ATrDO