Select an Analysis Mode
Select an Analysis ModeSelect an Analysis Mode
Select an Analysis Mode
5. Press the SELECT key to select a symbol, it will
blink. Select an analysis mode using the or
key.
If standard mode is used, select 5td .
If automatic mode is used, select U-a .
If manual mode is used, select U-m .
Set ACCURACY for Standard Mode
Set ACCURACY for Standard ModeSet ACCURACY for Standard Mode
Set ACCURACY for Standard Mode
6. Select ACCURACY with the SELECT key.
Select MID. with the or key.
When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating Pattern
Set the Heating PatternSet the Heating Pattern
Set the Heating Pattern
7. Select drying temperature with the SELECT key.
Select ramp drying of heating pattern with
the PROGRAM key.
Set the Final Drying Temperature
Set the Final Drying TemperatureSet the Final Drying Temperature
Set the Final Drying Temperature
8. Set 160°C with the or key.
Set the Ramp Time
Set the Ramp TimeSet the Ramp Time
Set the Ramp Time
9. Set 5.0 minutes with the or key.
Set the Preset Termination Value for
Set the Preset Termination Value for Set the Preset Termination Value for
Set the Preset Termination Value for Automatic
AutomaticAutomatic
Automatic
M
MM
Mode
odeode
ode
10. Select preset termination value with the SELECT
key. Select 0.05 [%/min] with the or key.
Set the
Set the Set the
Set the Measurement
Measurement Measurement
Measurement Unit
UnitUnit
Unit
11. Select measurement unit with the
SELECT key.
Select a moisture content (based on a wet
sample ) with the or key.
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
12. Select % display with the SELECT key.
Select 0.01 [%] with the or key.
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
13. Select gram display with the SELECT key.
Select 0.001 [g] with the or key.
Automatic
mode
0"%
"M"O"I"S"T
"/"W
"0."0"1%
"0."0"0"1g
0.000g
"1"6"0°C
ACCURAC
Y
MID.
"0."0"5
%/min
0.000g
5.0 min
0.000g
""1"0"5°C
Measurement unit
Minimum scale
value of % display
Minimum scale
value of gram
display
Final drying temperature
Ramp time
Ramp drying
ACCURACY
Preset termination value
Standard mode
Manual mode
"U"-"a
0.000g