A&D MF-50 Thermostat User Manual


 
MS-70 / MX-50 / MF-50 / ML-50 21
7.2.
7.2.7.2.
7.2. Quick Mode Operation
Quick Mode OperationQuick Mode Operation
Quick Mode Operation
The quick mode can obtain the moisture content with
settings of ACCURACY and drying temperature.
Sample is heated up at 200°C for approximately three
minutes so that moisture content is measured quickly.
7.2.1.
7.2.1.7.2.1.
7.2.1. ACCURACY
ACCURACYACCURACY
ACCURACY
ACCURACY of measurement can be set either HI, MID. or LO.
The sample quantity is automatically selected by ACCURACY.
The termination value of the analysis mode is automatically selected by
ACCURACY and minimum scale value of % display.
Analysis mode is the program to finish the measurement when a change of moisture
content per one minute becomes smaller than a preset termination value.
The settings are as follows: Specify an ACCURACY.
ACCURACY
Model Minimum scale HI MID. LO
0.001 % 0.02 %/min 0.05 %/min 0.05 %/min
0.01 % 0.05 %/min 0.10 %/min 0.20 %/min
MS-70
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.01 % 0.05 %/min 0.10 %/min 0.20 %/min
MX-50
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.05 %
0.1 %
MF-50
1 %
0.10 %/min 0.20 %/min 0.50 %/min
0.1 %
Preset
Termination
value
ML-50
1 %
0.20 %/min 0.50 %/min 1.00 %/min
Sample quantity 5 g 2 g 1 g
Use Precise result
Quick measurement
7.2.2.
7.2.2.7.2.2.
7.2.2. Operation
Operation Operation
Operation
This operation explanation uses the following example of the MX-50 :
Refer to "8.Selection of Measurement Method" for detail.
Input Parameters
Analysis mode ...................................Quick mode
Heating pattern ..................................Quick drying (
, Automatic selection)
Drying temperature............................ 130 °C
ACCURACY....................................... LO
Sample quantity.................................Approximately 1 g (Automatic selection)
Analyzing mode .................................0.20 %/min (Automatic selection)
Stored Parameters (Factory Settings for the MX-50)
Measurement unit ...............................Moisture content is based
on a wet sample
Minimum scale value of % display........0.01 %
Minimum scale value of gram display ... 0.001 g
Data memory function........................ Not used
% MOIST
D/W/D
200°C
Approx. 3 min.
Dr
y
in
g
tem
p
erature
Heating pattern