Tektronix TDS544A Water System User Manual


 
Nominal Traits
Specifications
Table 2-3: Nominal Traits — Time Base System
Name
Description
Range, Sample-Rate
1,3
TDS 540A, 544A
Number of
Channels On
1
2
3 or 4
Sample-Rate
Range
5 Samples/s – 1 GSamples/s
5 Samples/s – 500 MSamples/s
5 Samples/s – 250 MSamples/s
Range, Sample-Rate
1,3
TDS 520A, 524A
Number of
Channels On
1
2
Sample-Rate
Range
5 Samples/s – 500 MSamples/s
5 Samples/s – 250 MSamples/s
Range, Equivalent Time or Interpolated
Waveform Rate
2,3
500 MSamples/s to 100 GSamples/s
Range, Seconds/Division 500 ps/div to 10 s/div
Record Length
4
500 samples, 1000 samples, 2500 samples, 5000 samples,
15000 samples. A record length of 50000 samples is available with
Option 1M. Up to four 50 K waveform records may be saved in
NVRAM with Option 1M installed.
FastFrame
TM
Maximum Frame Rate: 50,000 Frames/Second
Frame Length Range: 50 points/Frame to 5,000 Points/Frame
Maximum Number of Frames:
910 Frames at 50 Points/Frame
(with Option 1M, 50,000 Record Length)
227 Frames at 50 Points/Frame
(standard configuration of 15,000 Record Length)
1 The range of real-time rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores the samples
in memory to produce a record of time-sequential samples
2 The range of waveform rates for equivalent time or interpolated waveform records.
3 The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by real-time sampling
of a single acquisition, the waveform rate is the same as the real-time sample rate; for a waveform created by interpolation of real-
time samples from a single acquisition or by equivalent-time sampling of multiple acquisitions, the waveform rate is faster than the
real time sample rate. For all three cases, the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveform
interval (WI) is the time between the samples in the waveform record.
4 The maximum record length of 15,000 samples (50,000 samples with Option 1M) is selectable with all acquisition modes except
Hi Res. In Hi Res, the maximum record length is 5,000 samples (15,000 samples with Option 1M).