Roper 4411-0070 Home Security System User Manual


 
16 NTE/CCD Detector Manual Version 2.A
To Mount the Lens on the Detector:
1.
Locate the large indicator dot on the side of the
lens.
2.
Note the corresponding dot on the front side of the
adapter.
3.
Line up the dots and slide the lens into the
adapter.
4.
Turn the lens counterclockwise until a click is
audible. The lens is now locked in place.
5.
In addition to the focusing ring of the lens, there is
provision for focusing the adapter itself. The
adjustment is secured by #4-40 setscrews on the
inside of the adapter. Directions for focusing the
lens and the adapter are provided in Chapter 4.
Lens release lever
Front part of adapter
for adjusting focus
Set screws to lock front
part of adapter in place
Figure 8. Nikon Lens Adapter
To Remove the Lens:
1.
Locate the lens release lever at the front of the lens mount.
2.
Press the lever toward the detector housing and simultaneously rotate the lens
clockwise.
3.
Then pull the lens straight out.
Although microscopes more commonly are used with a C-mount adapter, operation with
a detector having an F-mount adapter may also be possible. See Chapter 5, Microscopy
Applications and the adapter literature for further directions.
C-mount
NTE/CCD detectors can be ordered with an integral C-mount adapter. C-mount lenses
simply screw into the front of these detectors. Tighten the lens by hand only. See Chapter
5 for information on connecting to a microscope.
Note:
C-mount detectors are shipped with a dust cover lens installed. Although this lens
is capable of providing surprisingly good images, its throughput is low and the image
quality is not as good as can be obtained with a high-quality detector lens. Users should
replace the dust-cover lens with their own high-quality laboratory lens before making
measurements.
Overexposure Protection
Set the lens to the smallest aperture (highest F-number) and cover the lens with a lens
cap to prevent overexposure.