4-18 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 HFID
d. Preamp Assembly
1. Per Section 4-6, remove securing
hardware from electronics assembly
and lay on bench.
2. Disconnect and note location of ca-
bles.
3. Remove the two screws and washers
from the top bracket and slide the
preamp assembly out.
4. Remove the lower bracket from the
preamp assembly and install on re-
placement preamp assembly.
5. Slide replacement preamp assembly
into top bracket and secure with
mounting hardware.
6. Re-connect cables.
Figure 4-15. Preamp Assembly Installation
Bottom Bracket
Top Bracket
Preamp Assembly
656945