Emerson Process Management 760004-A Carbon Monoxide Alarm User Manual


 
Instruction Manual
760004-A
February 2002
4-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 HFID
Air Baffle (Ref)
102260
See Below
Sample Jet
657005
Gasket
102256
Jet Assembly
657012
Gasket
102273
Nut, Jet
657016
Air Baffle
.06
Position with tab over
the air hole
(approximately 1/16” to
the right).
End bent to raise it above the tab a
distance equal to the hei
g
ht of the tab.
7. Refer to Figure 4-8. Lift air baffle out
of burner base.
8. Remove the sample jet and gasket
from the bottom of the burner base.
9. Remove the jet nut. Grasp jet as-
sembly and lift out (along with upper
gasket) of burner base. Remove
bottom gasket.
Figure 4-8. Burner Jets