Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 2-9
Model NGA 2000 HFID
2-7 INSTALLATION CONSIDERATIONS CHECK-
LIST
Verify the following:
The Analyzer’s location should
be:
Clean
A well ventilated area
weatherproofed
Non-hazardous
Vibration-free
Have stable ambient temperature
The gas cylinders should be
equipped with a clean, hydrocar-
bon free two stage regulator and
shut off valve.
All external tubing, regulators,
valves, pumps, fittings, etc. are
clean.
The correct fuel type is being
used.
The THC content of the supply
gases are compatible with the
analysis range.
The calibration background
gases are similar to the sample.
The purge air out, burner ex-
haust, and bypass are vented to
atmospheric pressure. The pres-
sure should be constant.
The burner exhaust tube must be
slanted down a minimum of 6
degrees from horizontal.
The bypass line connection must
be slanted down a minimum of 6
degrees from horizontal for
drainage of water condensation.
If required, thermal insulation
around the bypass fitting to pre-
vent condensation in the bypass
restrictor.
If required, thermal insulation for
the sample inlet connection to
minimize the cold spot.
The heated line is at the correct
temperature.
The sample, zero, and span
gases are at the correct tem-
perature.
The heated line to have over
temperature protection.
The sample, bypass, and burner
exhaust tubing material must
handle high temperature and
have thermal insulation to protect
from burns.
The purge air out tubing to be
3/8 inch and less than 4 feet in
length.
All external gas connections
have been leak checked.
The dead volume for external
sample and fuel lines have been
minimized.
The stainless steel tubing used
for the fuel and sample lines is
clean.